Applications » Spectroscopy

Product code: Liquid Nitrogen Bath Cryostat

Overview

A range of Liquid Nitrogen bath cryostats with the sample either in vacuum or top loading into gas.

Range of windows

Suitable for the following range of spectroscopy techniques:
Raman
Infrared
Visible and ultra-violet spectroscopy
Photoluminescence
Photo-conductivity and electro-optical studies

  • Temperature range 77-800K
  • Stability ± 0.1K
  • Extended hold times
  • Window acces up to 60mm diameter

Product code: NVICEOPTI


TO CONFIGURE YOUR SYSTEM,
TALK TO US TODAY: 01993 706 444

System components available

Product code: Liquid Helium Variable Temperature Cryostat (with Optional Superconducting Magnet)

Overview

A helium bath cryostat with a variable temperature insert built in, allowing a choice of either dynamic, static, or sample in vacuum. Dimensions can be adjusted to increase the bath volumes. Ideal for use on beam lines, designed for use with VT probes, 3ICEVT systems and DICEVT systems.

Range of windows, see CRYObitz

Suitable for the following range of spectroscopy techniques:
Raman
Infrared
Visible and ultra-violet spectroscopy
Photoluminescence
Photo-conductivity and electro-optical studies
Visible and ultra-violet spectroscopy
Photoluminescence
Photo-conductivity and electro-optical studies

  • Temperature range 1.5 - 300 K
  • Stability ±5 K
  • Helium volume 8.0 ltrs
  • Nitrogen volume 11.0 ltrs
  • Window option 4 radial + 1 axial

Product code: VTICEBL


TO CONFIGURE YOUR SYSTEM,
TALK TO US TODAY: 01993 706 444

System components available

Product code: Liquid Helium Continuous Flow Cryostat

Overview

An adaptable modular range of static and dynamic continuous flow cryostats.

Sample holders, temperature range, tails dimensions length and diameter can all be modified to suit our customers needs

  • Single shot base temperature 1.2K
  • Helium consumption <1l/hour
  • Continuous flow temp 1.7K
  • Temperature range 1.2 - 800K

 Product code: CFICE


TO CONFIGURE YOUR SYSTEM,
TALK TO US TODAY: 01993 706 444

System components available

Product code: ESR system

Overview

Continuous flow helium cryostat specific designed for X-Band electron spin resonance spectroscopy. A range of cavities are available to suit all spectrometers

  • Temperature range from 1.9K to 325K
  • Temperature stability of +/- 100mK
  • Helium consumption of 1l/hr
  • Fast sample turn around times
  • Operate with Liquid Helium or Liquid Nitrogen

Product code: OPTI-ESRICECF

 


TO CONFIGURE YOUR SYSTEM,
TALK TO US TODAY: 01993 706 444

System components available

Product code: Top Loading Cryogen-free ³He (with optional superconducting magnet)

Overview

This top-loading cryogen-free 3He system, sample in liquid ³He, uses closed cycle cooler with vibration eliminated by a bellows system. A 1K stage allows full condensation of the ³He charge, this extends the hold time at a base temperature of 270mK to 70 hours, and gives a fast regeneration time. The temperature range for this system is 320mK to 300K temperature control across the range. Top-loading probe with 20mm sample space.

Options/customisation:

  • Further vibration isolation
  • Additional wiring
  • Coaxial cables
  • Wave guides
  • Line of sight ports
  • Optical windows
  • Base temperature: <270mK
  • Hold Time: 70 hours

Further additions:

  • Additional wiring
  • Coaxial cables
  • Wave guides
  • Rotator

 Product code: DRYICE3 TL


TO CONFIGURE YOUR SYSTEM,
TALK TO US TODAY: 01993 706 444

System components available