DICECold Probe
A compact cryogen-free ³He system for use in a UHV environment, sample in vacuum, uses a closed cycle cooler with vibration eliminated by a bellows system. A 1K stage allows full condensation of the ³He charge which extends the hold time at a base temperature of 300mK to 40 hours, and gives a fast regeneration time
The DICEDIPPER insert with a large diameter will be designed specifically to operate within the clear neck of a custom made vapour shielded helium dewar. The dipper will have a large sample space with room to mount various components to for experimental work. The height of the system will also be minimised where possible to ensure it will integrate into the purposed laboratory. The insert comes with a self-contained 3He gas charge and includes our patented condensation process.
The system can be combined with one of our anti-vibration frames to reduce the vibration at sample.
Temperature Range |
0.3K to 325K |
Cooling Power |
25µW / 50µW @ 325K (dependant on sample space) |
Temperature Stability |
<5mK below 1K |
Sample Environment |
Vacuum or Exchange Gas |
Sample Space |
50mm / 70mm |
Vibration at Sample |
|
Compressor |
Air or water cooled options available |
Wiring |
DC, RF and fiber optic options available |
Optical Access |
Up to five windows available (see window materials) |
Base Temperature |
<300mK with experimental wiring |
Hold Time |
>100 hours with experimental wiring, >70 hours in low vibration mode |
Temperature Controller |
Lakeshore 335 as standard (Other controllers available on request) |
Magnetic Field |
N/A |